G Model
CHROMA-353717; No. of Pages 7
ARTICLE IN PRESS
Journal of Chromatography A, xxx (2012) xxx–xxx
Contents lists available at SciVerse ScienceDirect
Journal of Chromatography A
journal homepage: www.elsevier.com/locate/chroma
Thermal Solid Sample Introduction–Fast Gas Chromatography–Low Flow Ion
Mobility Spectrometry as a field screening detection system
Saeed Hajialigol ∗ , Seyed Alireza Ghorashi, Amir Hossein Alinoori, Amir Torabpour, Mehdi Azimi
Iranian Space Agency, Engineering Research Institute, Engineering Research Center of Esfahan, Esfahan, Iran
a r t i c l e
i n f o
Article history:
Received 7 July 2012
Received in revised form 2 October 2012
Accepted 8 October 2012
Available online xxx
Keywords:
Gas Chromatography–Ion Mobility
Spectrometry (GC–IMS)
Thermal Solid Sample Introduction (TSSI)
Hyphenated technique
Field analysis
a b s t r a c t
The potential of Thermal Solid Sample Introduction (TSSI)–Fast Gas Chromatography (GC)–Low Flow Ion
Mobility Spectrometry (LF-IMS) having been designed and constructed in Engineering Research Center
of Esfahan, detector group was investigated for chemical detection capabilities. Customizing the configuration of fast GC–IMS as a high technology, provides unique solutions for rapid detection of a broad
range of chemical mixtures in many operational environments. TSSI configuration provides fast and easily applied method for direct detection with no additional sample preparation or extraction. The time
required for total analysis, less than 265 s, was determined by the wide range of solid matrixes, including
nitrate esters, nitroaromatics, and a nitramine. The fast extraction together with the short separation
time limits degradation of the thermally labile compounds and decreases the peak widths, which results
in larger peak intensities and a simultaneous improvement in detection limits. For signal-to-noise ratio
equals to 5, the detection limits for instrument for TNT, DNT and RDX were attained 15, 10 and 50 ng/l
respectively. The combination of short analysis time and low detection limits make this instrument a
potential candidate for field screening techniques.
© 2012 Elsevier B.V. All rights reserved.
1. Introduction
Ion Mobility Spectrometry (IMS) is a vanguard analytical technique that provides analytical information in a simple, rapid, and
inexpensive manner. For example, an ion mobility spectrometer can be used to extract global information about the volatile
profile of a sample [1,2]. Some advantages of IMS include the
following: low detection limits for single-component samples
(nanogram detection limits), fast analysis time (provides output
data in seconds), no required sample pretreatment of solids or analytes in solution, operation at atmospheric pressure eliminating
bulky vacuum systems, detection of both positive and negative ions
which provides good selectivity, capability to be miniaturized and
battery operated, and comparably less expensive to purchase and
operate than other trace detection technologies [3–9]. IMS instruments are the-state-of-the art for on-site detection of chemical
warfare agents, explosives and drugs [10] as well as air-quality
monitoring [11]. More recent areas of application have included
process control [12], rocket fuel leak detection [13], identification
of wood species [14], detection of bacteria [15], determining the
pesticides in agriculture products [16,17] and monitoring ammonia
∗ Corresponding author at: Iranian Space Agency, Engineering Research Institute,
Engineering Research Center of Esfahan, 7th Km Imam Khomeini Street, P.O. Box
81955/174, Esfahan, Iran. Tel.: +98 311 3222439; fax: +98 311 3222446.
E-mail address: s hajialigol@yahoo.com (S. Hajialigol).
levels in water [18]. IMS was used during space planned experiments for the Rosetta Comet Mission [19] and an IMS instrument
configured as a gas chromatography detector will be used to monitor cabin atmospheres in the next generation of space stations [20].
Thus, IMS has developed in recent years from few successful niche
applications to a general purpose analyzer suitable for a broad range
of applications.
The theory and operation of Ion Mobility Spectrometry has been
extensively described elsewhere [21]. While many IMS instruments
are used in the field to locate explosives, as with all detection
techniques, there are limitations. IMS does not handle chemical
mixtures adequately because of the complex interactions in the IMS
source that leads to obscure or less distinct plasmagrams [22–24].
IMS instruments are prone to inaccurate detection and false positive responses when chemical interferents are present in samples.
This technique is also relatively easy to saturate and the result is
a linear response range that is often limited to one to two orders
of magnitude. Thus, sample amount must be carefully controlled
at the instrument’s entrance to prevent saturation or nonlinear
response [23,25–27]. If these limitations can be mitigated, IMS can
be a more powerful tool in detecting explosives under a variety of
confounding sampling situations.
Coupling a gas chromatograph (GC) to the IMS detector is an
effective way to overcome IMS limitations [28–32]. While two
chemicals can have identical IMS ion mobilities, the chemicals typically have different GC retention times, which will be separated
prior to entering an IMS [33,34]. The addition of GC to IMS has
0021-9673/$ – see front matter © 2012 Elsevier B.V. All rights reserved.
http://dx.doi.org/10.1016/j.chroma.2012.10.010
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010
G Model
CHROMA-353717; No. of Pages 7
2
ARTICLE IN PRESS
S. Hajialigol et al. / J. Chromatogr. A xxx (2012) xxx–xxx
Fig. 1. Two designs for connecting capillary column to IMS (a) side connection and
(b) axial connection.
demonstrated improved trace organic chemical detection through
better resolution, reduced detector saturation, and increased IMS
linear range [35–38].
The modern age for mobility measurement after capillary gas
chromatography can be tracked back to the work of Baim and Hill
[28]. IMS was further demonstrated as a selective and nonselective
detector for GC in another report by Baim et al. when terpenes from
orange extracts were monitored for all product ions having drift
times between specific intervals [39]. The IMS constructed for interfacing to a GC was also investigated in negative ion mode [3]. The
detector was tuned to monitor chloride ion, and five pesticides at
100 pg cm−3 were investigated. Other studies that were carried out
with the constructed instrument included the positive ion mode
quantification of 2,4-dichlorophenoxyacetic acid residues in soil
extracts [40]. DeBono et al. developed GC–IMS technique for separation of different groups of mixtures such as narcotics, explosives,
lutidine isomers and pesticides [41,42]. Also in another study by
them, rapid analyses of pesticides on imported fruits were performed by GC–IMS [43].
An important aspect of the interface created for GC–IMS is the
transfer line. A number of prominent factors need to be considered
when developing a transfer line for an interface between a GC and
an IMS. The most important factors are maintaining the integrity
of the resolution, achieving quantitative transfers, and matching
flows [31].
There are two possible methods to connect a capillary column
to a mobility spectrometer including the attachment of the column
to the drift tube on axis to the ion source with a concentric inner
tube (Fig. 1b) or through the side of the drift tube (Fig. 1a).
The side design is effective, although complicated by the
requirements of a gas-tight connection between the column and
the drift tube. In the second design, the capillary column is extended
on axis into the ion source, from the end of the drift tube. In the axial
design, response is sensitive to the position of the capillary column
in the ion region. When a make up flow is used, the interface of the
GC to the IMS is uncomplicated and less sensitive to the location of
the capillary outlet, although the response is affected by the flow of
gas in the make up. The flow of the make up gas can be employed
to extend the linear range of response for a GC–IMS [21].
Achieving the minimal time of operation in gas chromatography (GC) has been a research topic ever since the introduction
of GC. Today, revived interest in fast GC is seen to be driven by
applications, such as process control and high-throughput analysis,
or by the desire to reduce the costs of operation and ownership in routine analysis. Having a newly designed oven with rapid
temperature programming capability (typically greater than 1 ◦ C/s)
and rapid cool down times on the order of 2 min, the fast GC instruments provides an effective injection system, advanced pneumatic
controls and fast data acquisition speeds. The net effect of these
instrumental designs is a GC analyses with analytical cycle times
in the range of 5 min or less. The goal of faster GC is to obtain the
information required from a certain sample in a shorter time. In
fast GC, the peaks elute in a shorter time frame and the registration system should be able to record and process the information
generated in real time, irrespective of which method of faster GC
has been used. In faster chromatography, more peaks are generated
per unit of time. This means that the peak broadening caused by the
detector must be small enough to preserve the column efficiency.
In addition, the sampling frequency of the detector must be high
enough to provide some 15–20 data points across the peak for an
accurate representation of the peak.
Complete characterization of a hazardous waste site can involve
a large amount of money, manpower and time. Sometimes, multimillion dollar analysis plans are carried out, only to find that
whether contamination levels are within standard guidelines. On
other occasions, a catastrophic environmental release requires
fast sample turnaround times to deal effectively with the release.
In these instances, classical methods of shipping samples to an
analytical laboratory are either a waste of resources or too time
consuming to provide timely data.
Idealistically, having the ability to identify and quantify analytes, a handheld, rugged field screening instrument should provide
instantaneous sample information with high signal to noise ratio
and resolution. Realistically, a functional field instrument needs
to meet some criteria including deploying by one person without extraneous equipment, being sufficiently sensitive to detect
compounds of interest, being capable of discriminating among the
various analytes (even with possible interferences present), providing a full analysis in less than 10 min and being rugged enough to
be moved about a sampling site without adverse effects. The need
for such an analytical instrument is considerable and its development would make the characterization of chemical contamination
extremely easier [44].
For increasing the capabilities of the GC–IMS as a field screening
instrument, a Thermal Solid Sample Introduction (TSSI) module is
needed for injection the real solid samples to gas chromatograph
without preparation of samples in laboratory. TSSI is suitable for
the fast and uncomplicated analysis of various analytes.
In this work a home-made fast GC was designed and constructed
for rapid separation of complex mixtures and minimum degradation of thermally labile compounds. Also a modified design of IMS
was made for interfacing it to capillary GC columns and rapid clear
down times. The performance of instrument was investigated by
injecting nitrate esters, nitroaromatics, and a nitramine individually and in complex mixtures through the constructed TSSI. Results
of experiments revealed the validity of TSSI–GC–IMS as an effective
instrument for field screening of explosives.
2. Instrumentation
TSSI–GC–IMS has the potential to satisfy the requisites for a
realistic and useful field screening instrument. TSSI–GC–IMS can
be divided in four main parts: Thermal Solid Sample Introduction,
Gas Chromatograph, Transfer line of GC to IMS and Ion Mobility
Spectrometer.
2.1. Thermal Solid Sample Introduction
TSSI as a simple, fast, precious and versatile sample preparation
technique works by vaporization for introducing a solid sample
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010
G Model
CHROMA-353717; No. of Pages 7
ARTICLE IN PRESS
S. Hajialigol et al. / J. Chromatogr. A xxx (2012) xxx–xxx
Fig. 2. Side view of TSSI assembly.
into gas chromatograph or gas detectors. It can be as a modern rapid alternative to complicated and laborious solid sample
preparation and injection technology with no solvent extraction
requirement for gas analyzers. TSSI is able to introduce separately
volatile, semi-volatile, and non-volatile organic compounds, from
the real solid sample matrices to fast response gas detectors. It
can introduce solid samples into analytical equipment in a manner that optimizes the concentration of the analyte of interest and
minimizes the concentration of interferences. This increases the
detector sensitivity for the analytes and lowers the detection limits. Advantages of TSSI versus solvent extraction include: no manual
sample preparation, no analytical interference from the solvent,
selective focusing/extraction of compounds of interest, cost effective reusable samplers and readily automated.
2.1.1. Principle of TSSI operation
The TSSI consists of five major parts: heated sample sweeper,
entrance slit, inlet of TSSI, six port valve and sample collecting loop
(Fig. 2). Samples are collected by suction with portable rechargeable
sampler onto porous mesh filter mounted in sample collector. To
begin an analysis the sample collector is inserted into the entrance
slit on the front of the TSSI–GC–IMS and start button is pressed.
When the analysis cycle starts, the heated sample sweeper moves
up and seals the sample collector to the inlet of TSSI. As the sample
is heated, the sample carrier gas sweeps the desorbed vapors into
a heated six port valve and cooled sample collecting loop where
it is trapped. For the sample analysis, the six port valve switches
position and the loop is rapidly heated. GC carrier gas purges the
loop and transfers the collected sample into the GC column. At the
end of analysis, the valve switches back into previous position and
the loop is then cooled for the next analysis.
2.1.2. Optimization of TSSI conditions
2.1.2.1. Decreasing the dead volume of desorption region and transfer
lines. Eliminating or reducing dead volume in TSSI is one of the
prominent aspects in designing and constructing this region.
The end of the TSSI is connected to gas chromatography capillary
column. Internal diameter of capillary columns varies between 0.25
and 0.53 mm. To decline the dead volume in this area, the internal
diameter of transfer lines and fittings need to be the same or near
the diameter of GC column.
This critical issue can be exacerbated in the inlet of TSSI which
has 10 mm diameter according to diameter of porous mesh filter.
The quantity of particles swept on the target collecting surface is
usually very small, often in the range of only nanograms or even
picograms per square centimeter. Inasmuch as the porous mesh
filter should collect maximum amounts of sample particles from
environment and transfer them to TSSI for analyzing, it needs to be
provided with diameter at least 10 mm. On the other side, mesh filters having bigger diameters would absorb the temperature of the
desorption region, the space between heated sample sweeper and
3
inlet of TSSI, and cause a salient drop in its temperature. Thus, the
particles collected on the filter cannot evaporate considerably and
a few amounts of them will be delivered to the instrument. With
regard to the aforementioned remarks, the inlet diameter of TSSI
should be equal to 10 mm similar to the diameter of the porous
mesh filter. Moreover, to fit this diameter with the required one
(about 0.5 mm), the multi-stage decrease should be implemented.
It goes without saying that if we perform sudden, unorganized
reduction, we will produce some steps in the gas transfer lines
causing noticeable dead volumes. On the contrary, by decreasing
the diameter in various stages as well as arranging cone-shaped
structures, we overcome this problem effectively.
The end of the TSSI was connected to gas chromatography capillary column. Internal diameter of capillary columns varies between
0.25 and 0.53 mm. A key factor in the performance of coupling
with capillary GC is the absence of any dead volume. The term
‘dead-volume’ refers to volumes within the chromatographic system which are not swept by the mobile phase. By dead volume we
mean the empty spaces in the TSSI and connecting tubing before the
column and the connecting tubing and ion mobility spectrometer
after the column. In these sections, band broadening occurs without realizing separation of components. Also dead volume causes
to loss some of analytes during transferring to GC capillary column
and IMS. It results loss of efficiency to the chromatographic system and increasing detection limit [45]. Eliminating or reducing
dead volume in TSSI is one of the important aspects of design and
construction of TSSI.
For reducing dead volume, internal diameter of transfer lines
and fittings should be equal or near to internal diameter of GC capillary column. On the other side, this critical problem can be found
in the inlet of TSSI. TSSI instrument works by releasing analytes
from a solid sample collected on porous mesh filter by thermally
desorbing in desorption region and then transferring into inlet of
TSSI which dead volume is reduced in it.
2.1.2.2. Decreasing the sample loss in desorption region. The instrument should provide a signal that is proportional to the
concentration of target molecule vapor. This concentration is
further dependent on the equilibrium vapor pressure of target
molecule, the vaporization temperature, the total flow rate of
non-target gas that dilutes the target vapor and sample losses on
desorption region of the TSSI.
Sample losses after evaporation of samples in desorption region
is one of the major problems in TSSI. During the vaporization process of a liquid or solid sample, the pressure drastically increases in
desorption region and causes to lose vapor sample from this part. To
overcome this problem, the pressure of this region was reduced by
utilizing a sampling pump which sucks the sample gas into the loop
from the inlet of TSSI. Also for decreasing sample loss in desorption region during the vaporization process, this region should be
sealed. Because of this fact, the material of sample collector was
optimized. A sample collector consists of two parts, a porous mesh
filter and PTFE card. Due to the flexibility of PTFE in high pressure and temperature, it can be used as a suitable seal material for
sample collector.
2.2. Fast Gas Chromatography
A home-made fast gas chromatographic system was designed
and built to have minimum temperature gradients. The air within
the oven is heated with an electric heating element placed within
the oven and circulated with a fan to uniformly heat or cool the interior of the oven and the column. Generally, ovens have a rectangular
or square interior. The square or rectangular shape of the oven interior makes it difficult to uniformly heat the oven. Also heating time
is increased as is the cooling time, because of large thermal mass of
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010
G Model
CHROMA-353717; No. of Pages 7
4
ARTICLE IN PRESS
S. Hajialigol et al. / J. Chromatogr. A xxx (2012) xxx–xxx
responses, a cylindrical make up gas was flown into the interface
around the analyte input tube.
In brief, with implementing the mentioned method, the advantages of this configuration would be gained including preventing
from detector saturation of the instrument, rapid clear down times,
extend the linear dynamic range of the instrument, control ion
chemistry by maintaining the concentrations of analyte in the
source region and high efficiency of dispersion of analyte in reaction
region.
2.4. Low flow ion mobility spectrometer
Fig. 3. Cross section of cylindrical oven of fast GC.
the oven interior. In the TSSI–GC–IMS, the oven cavity is cylindrical
(Fig. 3), thereby reducing the surface area and volume of the cavity
as compared to square or rectangular oven cavities with removing corner tips, further decreasing the heat load and therefore the
heating and cooling time.
Fast GC of this instrument provides fast and linear heating rate
up to 70 ◦ C/min. Temperature gradient is less than 1 ◦ C for 40 ◦ C/min
and around 2 ◦ C for 70 ◦ C/min heating slopes. This allows determining a large number of compounds with wide range of boiling points.
Also applying a shorter column length, an optimum carrier gas
velocity and faster temperature programming cause to determine
temperature labile compounds too.
On the other hand, detection system should be sufficiently fast
and sensitive. The constructed IMS along with its sophisticated data
acquisition system indicated its potential as a detector for fast GC.
Baseline width of our peaks in IMS is about 1.2 ms, this requires a
sample frequency of 12.5–16.7 kHz to provide 15–20 data points
across the peak for an accurate representation. Typical spectral
acquisition rate of our IMS is 25 spectra/s, each spectrum consists
of 4096 samples collected in 40 ms. These figures implied that the
sampling frequency of the system is about 100 kHz which provides
our IMS as a well qualified detector system to couple with a GC to
form a capable fast GC–IMS.
2.3. Transfer lines of GC to IMS
The most important factors are maintaining the integrity of the
resolution, achieving quantitative transfers, and matching flows
[32].
There are two possible methods to connect a capillary column
to a mobility spectrometer including the attachment of the column
to the drift tube on axis to the ion source with a concentric inner
tube (Fig. 1b) or through the side of the drift tube (Fig. 1a).
The flow of the make up gas can be employed to extend the
linear range of response for a GC/IMS [46].
Make up gas allows vapors eluting from the fused-silica column to be dispersed into the reaction region effectively because
sample and reactant ions are mixed well. When the make up flow
is reduced, effluent from the fused-silica column swept along the
walls of tubing declines, and the mixing of reactant ions and sample is reduced. Consequently, response is variable, depending upon
make up flows.
In the constructed TSSI–GC–IMS, the axial design with unidirectional flow was used for connection of GC column and IMS.
A gas chromatography–ion mobility spectrometer inlet based on
this configuration causes to rapid clear down times, an important feature for gas chromatography detectors. In order to improve
The low flow IMS was designed and constructed at Engineering
Research Center of Esfahan (Esfahan, Iran). The developed configuration constitutes an open-system easy to manufacture at low cost.
The device comprises the parts of a regular ion mobility spectrometer, namely, an ionization chamber, shutter-grids, a separation
chamber, an ion collector and voltage generators. The device is
equipped with a radioactive 63 Ni ionization source (St. Petersburg,
Russia), with an activity of 12 mCi which consists of a cylindrical
cavity of 1 cm in length and 1 cm in diameter houses the Ni-foil.
A sample inlet lets a continuous stream of the gas carrying the
analytes passing through the ionization chamber, where the ions
are effectively focused before entering the separation chamber. To
transport ions from the ionization chamber into the drift tube, a
shutter-grid built from parallel steel wires is used. The configuration type of the shutter follows a Bradbury–Nielsen gate design.
Electrical components allow adjusting the opening time of the grid
to let the ions pass at short pulses towards the separation chamber. The drift tube, with an inner diameter of 15 mm and 75 mm
in length, was designed on the basis of avoiding ion repulsion and
charge effects and allowing homogeneity of the drift field. In this
separation chamber, kept at 120 ◦ C, ions with different mobility are
separated within an electrical field of 280 V/cm at a fixed voltage of
3 kV hold by means of 10 parallel drift rings constructed from steel
316 which are coaxial cylinders.
A high voltage supply is connected to the drift rings placed in
equal distance to create the electric field in the drift tube. Isolators
made from Teflon and resistors are placed between the drift rings.
The influence of the voltage in the signal and separation of the ions
was characterized using acetone as exemplary substance for IMS
separation. The optimized value is 3 kV as a compromise between
sensitivity and the better resolution. The drift gas (nitrogen) flows
in the opposite direction of the ions in order to prevent non-ionized
impurities caused by water molecules or volatile components of
the column stationary phase, from entering the separation chamber. A gas inlet at the front of the drift tube allows the entrance of
the carrier gas and an outlet line makes the gaseous streams exit
the instrument. Finally, the separated ions reach a Faraday plate
of 35 mm in diameter and the signal is delivered to the computer.
Home-made data acquisition and analysis software was used to
acquire the signal of ions collected by the Faraday plate. The modification on conventional IMS was done to improve the sensitivity and
use as a GC detector. In this manner, the ionization cell volume was
reduced. The reduction of ionization cell volume decreases the residence time of a compound in the ionization region to a few tenths
of a second, maintaining the integrity of the gas chromatographic
separation and improve sensitivity. Also this modification causes a
decrease in gas consumption of IMS. Similar to uni-directional flow
design in transfer line, reduction of ionization cell volume causes
rapid clear down time of IMS. Consequently the consumed drift gas
would be reduced twenty times compared to the volume of drift gas
in conventional home-made IMS. According to Table 1 sum of drift
gas and make up gas was reduced from 550 ml/min to 25 ml/min
in low flow IMS.
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010
ARTICLE IN PRESS
G Model
CHROMA-353717; No. of Pages 7
S. Hajialigol et al. / J. Chromatogr. A xxx (2012) xxx–xxx
Table 1
Comparison of the characteristics of the low flow ion mobility spectrometer and
conventional IMS.
Parameter
Low flow IMS
Traditional IMS
Ionization source
Length of the drift region
Full length of the spectrometer
Electrical field strength
Make up flow
Drift flow
63
63
Ni (12 mC)
75 mm
110 mm
280 V/cm
10 ml/min
15 ml/min
Ni (16 mC)
120 mm
230 mm
240 V/cm
250 ml/min
300 ml/min
Comparison of the characteristics of both detectors (conventional and low flow IMS) is summarized in Table 1.
5
Table 2
TSSI–GC–IMS experimental conditions.
Conditions
IMS operating mode
Sampling time (s)
TSSI temperature (◦ C)
Drift tube temperature (◦ C)
Loop (preconcentrator) temperature (◦ C)
Valve temperature (◦ C)
Oven initial temperature (◦ C)
Oven initial hold (s)
Oven ramp rate (◦ C/min)
Oven final temperature (◦ C)
Final hold (s)
Transfer line temperature (◦ C)
Total analysis time (sampling + analyzing)
Negative
60
235
110
240
220
140
30
25
200
30
220
264
4. Results and discussion
3. Materials and methods
4.1. TSSI–GC–IMS analysis of individual explosives
All reagents were of analytical grade. Acetone was purchased from Merck. Trinitrotoluene, dinitrotoluene and
1,3,5-trinitroperhydro-1,3,5-triazine were purchased from
Sigma–Aldrich Chemical Company (St. Louis, MO, USA). The
standards were weighed into 2 ml amber vials and were dissolved
with 1 ml of acetone, and then the vials were placed into an
ultrasonic bath for effective dissolution for 10 min then diluted to
10 ml for making stock solutions. Other solutions were made by
successive dilution of the stock solutions.
In the TSSI–GC–IMS, analytes are thermally desorbed in a Thermal Solid Sample Introduction (TSSI). Then desorbed analytes enter
a sample loop which acts as a preconcentrator for GC separation.
The sample loop (preconcentrator) is subsequently heated and
sample vapors are carried into the GC column (15 m DB-1; 0.53 mm
internal diameter; 1.0 m film thickness). Experimental conditions
for each of the three explosives analyzed are listed in Table 2.
To accurately transfer a known quantity of explosive to each
swab, 1 l of a dissolved explosives solution was directly deposited
onto sample collector. The solvent was allowed to evaporate (5 s),
leaving the desired mass of explosive on the swab before it was
placed into the TSSI.
The TSSI–GC–IMS chromatogram of 40 ng TNT is shown in
Fig. 4a. The spectrum shows an intense peak at a retention time
of 162 s. The response of instrument to DNT is more interesting
due to the fact that the TSSI–GC–IMS indicates its potential to separate two isomers of DNT. The results of introducing 80 ng DNT has
been shown in Fig. 4b in which two distinct peaks can be observed,
a sharp peak with retention time of 140 s and a small one with
shorter retention time of 135 s related to 2,4-DNT and 2,6-DNT
respectively. Eventually, the chromatogram of 100 ng RDX has been
revealed in Fig. 4c. Although this figure demonstrates a keen peak
with retention time of 80 s which is assigned to byproducts of RDX,
the intrinsic peak of this compound can be found in 200 s.
4.2. Identification of interested compounds in mixture sample
In this study, 1 l standard mixed solutions of 40 ppm
2,4,6-trinitrotoluene, 50 ppm 2,4-dinitrotoluene and 100 ppm
1,3,5-trinitroperhydro-1,3,5-triazine were directly deposited onto
sample collector. The solvent was allowed to evaporate leaving the
desired mass load of explosives on the sample collector before it
Fig. 4. Chromatogram of (a) TNT, (b) DNT, (c) RDX and (d) mixture of explosives.
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010
G Model
CHROMA-353717; No. of Pages 7
6
ARTICLE IN PRESS
S. Hajialigol et al. / J. Chromatogr. A xxx (2012) xxx–xxx
was placed into the TSSI. The chromatogram of these compounds
has been shown in Fig. 4d.
Recent findings reveal that the designed TSSI–GC–IMS has been
well qualified to separate and identify each of the compounds in
the complicated matrix.
4.3. Detection limits
In TSSI–GC–IMS, the instrument is designed to pass sample and
carrier gas for a preset period of time through a preconcentrator
described earlier which essentially acts as a trapping medium. Subsequently, a six-port valve reverses flow and the preconcentrator
is simultaneously heated to desorb and transfer the analytes into
the GC column for chemical separation. All preconcentrators have
a limited trapping capacity, which can differ among compounds.
Once a preconcentrator reaches its maximum trapping capacity for
an analyte, any subsequent analyte will simply vent and may never
be transferred into the GC column or, in this case, the IMS.
The TSSI–GC–IMS response in comparison with the stand alone
IMS response has lower peak amplitude at nearly all concentrations. The decreased signal response for TSSI–GC–IMS at the same
concentrations may be primarily attributed to the limiting effect
of the sample loop, since analyte is lost when the preconcentrator
reaches its maximum trapping capacity. The decreased response
may also happen due to inefficient transfer of sample vapors into
the GC or the IMS.
When samples are provided as discrete items, such as filter or swipes or as solid-phase microextraction fibers, MDLs are
expressed as picograms per sample or nanograms per sample. This
is commonly practiced with semivolatile compounds and when IMS
drift tubes are used as chromatographic detectors.
Detection limits, defined as the fivefold signal-to-noise ratio,
were attained 15, 10 and 50 ng respectively. They were higher than
IMS alone.
It is worth mentioning that TSSI–GC–IMS, unlike Mass Spectrometry, operates at ambient pressure and for this reason the
detection limit of IMS is higher than MS. On the other hand, in IMS
only a limited amount of materials are introduced into the drift tube
that depends on the time in which the shutter grid is open. Additionally, MS spectrometry has a more powerful ionization source
compared to IMS. These reasons exacerbate the detection limit of
IMS in contrast with MS. The limits of detection for the GC–MS
analyses of TNT and RDX using the Rtx5MS column (15 m length,
0.53 mm inner diameter, 1.5 m film thickness) have been acquired
6 and 3 ng/l respectively [45].
4.4. Linear dynamic range
The linear dynamic range (LDR) of the instrument was evaluated. Response curves are shown in Fig. 5 for the three selected
explosive compounds. Inasmuch as TSSI–GC–IMS limits the amount
of sample entering the IMS, no saturation will happen for the IMS
detector and thereby the instrument shows its reliability to have a
wide linear dynamic range.
Fig. 5a shows a linear range for DNT from 10 to 200 ng which is
followed by a second linear region from 200 to 1500 ng, where the
ion source is nearly saturated. Thus, the dynamic or working range
for DNT is from 10 to 1500 ng. The practical tests for TNT revealed
a greater dynamic range, 15–2500 ng; a linear range from 15 to
150 ng and another range from 150 to 2500 ng (Fig. 5b). Albeit TNT,
in comparison with DNT, has a narrower dynamic range for low
concentrations, it shows a large range for higher ones which is well
enough to provide a wider dynamic range in general. The achieved
range for RDX is 50–5000 ng including a linear range from 50 to 500
and a second one from 500 to 5000 ng (Fig. 5c).
Fig. 5. Response curves for (a) DNT, (b) TNT, and (c) RDX.
Table 3
Reproducibility of TSSI–GC–IMS for various explosive.
TNT
DNT
RDX
Precision%
Reproducibility%
5.34
10.43
17.32
12.04
13.41
15.23
The calibration curves of the compounds showed excellent
linearity over a concentration range of at least two orders of magnitude having r2 -values greater than 0.97. These results are especially
promising with respect to the need of establishing a quantitative
IMS determination method for explosives.
With analysis of DNT, TNT and RDX, it appears that the instrument has a greater LDR over that of IMS alone [46]. Commonly,
linear ranges of 10–100 have been reported for IMS [21].
4.5. Precision and reproducibility study of TSSI–GC–IMS
Six replicate of 100 ng of the standard solution were analyzed
to determine the precision of the instrument. The results revealed
that RSDs were less than 18% for all explosives (Table 3).
Also in order to assess the reproducibility of the Instrument
100 ng of each solution was analyzed in triplicate on four days
during in one month and the RSD for each analyte was determined.
It is distinct that TSSI–GC–IMS produces inexhaustible and reliable results for each compound during analysis in one month.
5. Conclusion
A TSSI–GC–IMS, suitable for use in explosive field screening
detection applications, was designed and constructed. This
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010
G Model
CHROMA-353717; No. of Pages 7
ARTICLE IN PRESS
S. Hajialigol et al. / J. Chromatogr. A xxx (2012) xxx–xxx
instrument has the advantages of a simple, rugged system along
with quick response and fast clear down time. It was shown to be
selective and sensitive which can be applied to identification of
interested analyte in complex mixtures. Results indicate excellent
linearity over concentration ranges and reproducibility that make
it reasonable for quantitative analysis as well. The next goal is to
make the instrument suitable for space applications and utilize it as
a powerful and sensitive detector in ground laboratories and inside
the spacecraft cabins.
References
[1] R. Garrido-Delgado, F. Mercader-Trejo, S. Sielemann, W. de Bruyn, L. Arce, M.
Valcarcel, Anal. Chim. Acta 696 (2011) 108.
[2] N.S. Chary, A.R. Fernandez-Alba, Trends Anal. Chem. 32 (2012) 60.
[3] M.A. Baim, H.H. Hill Jr., J. High. Resolut. Chromatogr. Chromatogr. Commun. 6
(1983) 4.
[4] F.W. Karasek, W.D. Kilpatrick, M.J. Cohen, Anal. Chem. 43 (1971) 1441.
[5] J.I. Baumbach, G.A. Eiceman, Appl. Spectrosc. 53 (1999) 338A.
[6] G.A. Eiceman, J.A. Stone, Anal. Chem. 76 (2004) 390A.
[7] M. Utriainen, E. Krpnoja, H. Paakkanen, Sens. Actuators B: Chem. 93 (2003) 17.
[8] D. Li, M. Tam, P. Dutta, H.H. Hill Jr., Encyclopedia of Microfluidics and Nanofluidics, Springer, USA, 2008, pp. 1368–1376.
[9] S. Zimmermann, S. Barth, IEEE, miniaturized ion mobility spectrometer for
detection of hazardous compounds in air, in: Transducers ’07 & Eurosensors
Xxi, Digest of Technical Papers, vols. 1 and 2, 2007, pp. U759–U760.
[10] Y. Seto, Yakugaku Zasshi, J. Pharm. Soc. Jpn. 126 (2006) 1279.
[11] G.A. Eiceman, S. Sowa, S. Lin, S.E. Bell, J. Hazard. Mater. 43 (1995) 13.
[12] F. Li, Z. Xie, H. Schmidt, S. Sielemann, J.I. Baumbach, Spectrochim. Acta B: Atom.
Spectrosc. 57 (2002) 1563.
[13] G.A. Eiceman, M.R. Salazar, M.R. Rodriguez, T.F. Limero, S.W. Beck, J.H. Cross, R.
Young, J.T. James, Anal. Chem. 65 (1993) 1696.
[14] A.H. Lawrence, R.J. Barbour, R. Sutcliffe, Anal. Chem. 63 (1991) 1217.
[15] A.P. Snyder, D.B. Shoff, G.A. Eiceman, D.A. Blyth, J.A. Parsons, Anal. Chem. 63
(1991) 526.
[16] K. Tuovinen, H. Paakkanen, O. Hnninen, Anal. Chim. Acta 404 (2000) 7.
[17] M.T. Jafari, M. Azimi, Anal. Lett. 39 (2006) 2061.
[18] A.R.M. Przybylko, C.L.P. Thomas, P.J. Anstice, P.R. Fielden, J. Brokenshire, F. Irons,
Anal. Chim. Acta 311 (1995) 77.
[19] D.R. Kojiro, G.C. Carle, D.E. Humphry, M. Shao, N. Takeuchi, in: The Fourth
International Workshop on Ion Mobility Spectrometry, Cambridge, UK,
1995.
7
[20] T. Limero, A. Brittain, J. Breach, in: The Sixth International Workshop on Ion
Mobility Spectrometry, Bastei, Germany, 1997.
[21] G.A. Eiceman, Z. Karpas, Ion Mobility Spectrometry, second ed., CRC Press, Boca
Raton, FL, 2005.
[22] L.M. Fytche, M. Hupe, J.B. Kovar, P. Pilon, J. Forensic Sci. 37 (1992) 1550.
[23] G.A. Eiceman, W. Yuan-Feng, L. Garcia-Gonzalez, C.S. Harden, D.B. Shoff, Anal.
Chim. Acta 306 (1995) 21.
[24] G.E. Spangler, J.P. Carrico, D.N. Campbell, J. Test. Eval. 13 (1985) 234.
[25] H.H. Hill, W.F. Siems, R.H. St. Louis, Anal. Chem. 62 (1990) 1201A.
[26] A.P. Snyder, C.S. Harden, A.H. Brittain, M.G. Kim, N.S. Arnold, H.L.C. Meuzelaar,
Anal. Chem. 65 (1993) 299.
[27] L.M. Matz, P.S. Tornatore, H.H. Hill, Talanta 54 (2001) 171.
[28] M.A. Baim, H.H. Hill, Anal. Chem. 54 (1982) 38.
[29] A.P. Snyder, W.M. Maswadeh, J.A. Parsons, A. Tripathi, H.L.C. Meuzelaar, J.P.
Dworzanski, M.G. Kim, Environmental Monitoring and Remediation Technologies II, 1999, p. 122.
[30] Y. Lu, P.B. Harrington, Anal. Chem. 79 (2007) 6752.
[31] A.B. Kanu, H.H. Hill Jr., J. Chromatogr. A 1177 (2008) 12.
[32] G.W. Cook, P.T. LaPuma, G.L. Hook, B.A. Eckenrode, J. Forensic Sci. 55 (2010)
1582.
[33] M. Juenger, B. Boedeker, J.I. Baumbach, Anal. Bioanal. Chem. 396
(2009) 471.
[34] C. Kwan, A.P. Snyder, R.P. Erickson, P.A. Smith, W.M. Maswadeh, B. Ayhan, J.L.
Jensen, J.O. Jensen, A. Tripathi, IEEE Sens. J. 10 (2010) 451.
[35] K.J. Budde, W.J. Holzapfel, M.M. Beyer, S.C.I. Inst Environm, IES 1993
Proceedings, Contamination Control: Symposium on Minienvironments; Symposium on Biocontamination Control, vol. 1, 1993, p. 366.
[36] E. Aguilera-Herrador, S. Cardenas, V. Ruzsanyi, S. Sielemann, M. Valcarcel, J.
Chromatogr. A 1214 (2008) 143.
[37] M.A. Baim, R.L. Eatherton, H.H. Hill, Anal. Chem. 55 (1983) 1761.
[38] Z. Karpas, Y.-F. Wang, G.A. Eiceman, Anal. Chim. Acta 282 (1993) 19.
[39] M.A. Baim, F.J. Schuetze, J.M. Frame, H.H. Hill, Am. Lab. 14 (1982) 59.
[40] M.A. Baim, H.H. Hill Jr., J. Chromatogr. 279 (1983) 631.
[41] R. DeBono, Am. Lab. 33 (2001) 26.
[42] R. DeBono, A. Grigoriev, R. Jackson, R. James, F. Kuja, A. Loveless, A. Nacson, A.
Rudolph, Y. Sun, J. Int, Ion Mobility Spectrom. 5 (2002) 194.
[43] R. DeBono, A. Grigoriev, R. Jackson, R. James, F. Kuja, A. Loveless, A. Nacson, A.
Rudolph, Y. Sun, J. Int, Ion Mobility Spectrom. 4 (2001) 16.
[44] D.A. Atkinson, Evaluation of Gas Chromatography Coupled with Ion Mobility
Spectrometry for Use as a Field Screening Tool, Washington State University,
Washington, 1994.
[45] J.A. Mathis, Development of Chromatography and Mass Spectrometry Methods
for Explosives Analysis, The College of Art and Science, Ohio University, 2004,
pp. 115–116.
[46] M.A. Baim, H.H. Hill, J. Chromatogr. 299 (1984) 309.
Please cite this article in press as: S. Hajialigol, et al., J. Chromatogr. A (2012), http://dx.doi.org/10.1016/j.chroma.2012.10.010